Abstract
This work aims to establish some X-ray qualities recommended by the International Standard Organization (ISO) using the half-value layer (HVL) and Hp(10) dosimetry approaches. The HVL values of the following qualities N-60, N-80, N-100, N-150 and N-250 were determined using various attenuation layers. The obtained results were compared to those of reference X-ray beam qualities and a good agreement was found (difference less than 5% for all qualities). The GAMOS (Geant4-based Architecture for Medicine-Oriented Simulations) radiation transport Monte Carlo toolkit was employed to simulate the production of X-ray spectra. The characteristics HVLs, mean energy and the spectral resolution of simulated spectra have been calculated and turned out to be conform to the ISO reference ones (difference less than the limit allowed by ISO). Furthermore, the conversion coefficients from air kerma to personal dose equivalent for simulated and measured spectra were fairly similar (the maximum difference less than 4.2%).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.