Abstract

The aim of this study was to test C/C material (carbonized, graphitized or covered with pyrolytic carbon) designated for the use in orthopaedic and bone surgery. Using an in vitro assay we confirmed, that the cell proliferation was exhibited the mostly on the C/C composite coated with pyrolitic carbon and afterwards polished. The two latest of subsequent water extracts of this material had a slightly inhibiting effect on the cells metabolic activity. Biocompatibility test in vivo performed subcutaneously on rats did not show big differences between three tested implants (C/C composite, epoxy resine, titanium alloy), on the other hand the plates tested on pigs demonstrated foreign-body reaction induced by wear C/C composite material. Such debris were found both in the neighborhood of the implant as well as in the lymphatic node.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.