Abstract

A three-dimensional nonlinear finite element model (FEM) was developed for a parametric study that examined the effect of synthetic augmentation on nonfractured vertebrae. The objective was to isolate those parameters primarily responsible for the effectiveness of the procedure; bone cement volume and bone density were expected to be highly important. Injection of bone cement was simulated in the FEM of a vertebral body that included a cellular model for the trabecular core. The addition of 10% and 20% cement by volume resulted in an increase in failure load, and the larger volume resulted in an increase in stiffness for the vertebral body. Placement of cement within the vertebral body was not as critical a parameter as cement amount. Simulated models of very poor bone quality saw the best therapeutic benefits.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.