Abstract

Bacillus subtilis var. niger spores were used to determine the exposure time for formaldehyde decontamination of biological safety cabinets. Formaldehyde contact times less than 3 hr were insufficient for sterilization. A contact time of 4 hr or more resulted in a reproducible killing of the spore strips placed inside the cabinets. At 6 hr sufficient formaldehyde had diffused through the high efficiency particulate air (HEPA) filter to sterilize the strips with lower spore counts. A minimum of 5 to 6 logs of killing were observed after 4 to 6 hr of treatment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.