Abstract

Oxygen-sensing implants, composed of paramagnetic microcrystals embedded in a biocompatible polymer, are increasingly being used for electron paramagnetic resonance (EPR) oximetry in animal models and human subjects. The implants are stable and designed to stay in the tissues for indefinite periods. However, it is not known whether the crystals that may be exposed on the surface of the implants or leached out from the implants will induce cytotoxicity thereby compromising their biocompatibility over the long term. The goal of the current study was to evaluate the cytotoxicity of the implants and crystalline particulates under in vitro conditions. Apoptosis and cell viability studies were performed using L6, a rat muscle cell line and AsPC-1, a cancer cell line derived from human pancreatic adenocarcinoma. The results indicated that neither the intact implants nor their components elicit cytotoxicity, thus establishing their biocompatibility for use in human subjects.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.