Abstract

Background: Volar locking plate (VP) and Kirschner wire (K-wire) fixations of distal end radius fractures are the most frequently used techniques that produce similar long term clinical results. However, inadequate fixation strength of the K-wire may cause pin loosening or migration. Although these complications can be prevented by immobilization, joint stiffness and a prolonged recovery period can occur.
 Objective: Herein, a technique that provided more stability, allowing immediate motion after fixation by linking the K-wires into a single system (locked K-wire system) was proposed.
 Methods: We evaluated biomechanical responses of the locked K-wire system and a VP in extraarticular distal radius fracture models AO/OTAa type 23A2 and 23A3 using three-dimensional finite element analysis. All models were tested under axial, bending, and torsional loads.
 Results: From the simulation results, the total displacement was greater in the dorsal wedge fracture than that from the simple fracture under all loads for both fixation systems. The locked K-wire system and the VP could withstand immediate physiologic load with maximum displacements of 1.15 mm and 1.39 mm, respectively.
 Conclusion: Considering the immediate physiologic load resistance and the ability to preserve its position during the bone-healing period, the locked K-wire system might be used as an alternative to fix distal radius fractures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.