Abstract

Endoscopic imaging in tubular structures, such as the tracheobronchial tree, could benefit from imaging optics with an extended depth of focus (DOF) to accommodate the varying sizes of tubular structures across patients and along the tree within the same patient. Yet the extended DOF needs to be accomplished without sacrificing resolution while maintaining sufficient sensitivity and speed of imaging. In this Letter, we report on the measured resolution and sensitivity achieved with a custom-made micro-optic axicon lens designed to theoretically achieve an 8 mm DOF. A measured invariant resolution of approximately 8 microm is demonstrated across a 4 mm measured DOF using the micro-optic axicon while achieving an invariant sensitivity of approximately 80 dB with a 25 mW input power. Double-pass Bessel beam spectral-domain optical coherence tomography with an axicon micro-optic lens (i.e., <1 mm in diameter) is, for the first time to our knowledge, demonstrated in a biological sample demonstrating invariant resolution and signal-to-noise ratio across a 4 mm measured DOF, which is compared to Gaussian beam imaging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.