This study identified a novel method of validating the force and Centre of Pressure (COP) obtained from pressure-measuring insoles, utilizing commercially available equipment. The focus of this research was to compare force and COP data between a newly developed Smart Insole (SI) and the Pedar (PE) insole using an independent point of reference device – the Kistler force plate (FP). Results showed good correlation between forces measured from both systems and the FP (R2 > 0.98) with RMSE of 11.2% and 5.6% respectively. For the COP, the Systematic Error (SE), k, and R2, are presented. For the SI, the COPx and COPy data achieved k = 0.763 and 1.009 (R2 = 0.855 and 0.990). For the PE, k = 0.155 and 1.0143 (R2 = 0.014 and 0.995). The SI is comparable to PE in terms of accuracy when measuring force and COP, with the advantages of being accurate, portable and cheaper.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.