Abstract

AbstractWith the increasing use of ion‐selective electrodes in environmental and health applications, it is important to know the precision of estimated concentrations. A Bayesian model for non‐linear calibration is introduced which provides estimates of measurement precision by incorporating uncertainty in calibration parameters and inherent random noise in emf response. The analysis of lead in 17 soil samples demonstrates that large gains in precision are possible when calibrations are extended to include multiple electrodes and standard addition data. The results highlight the need for improved calibration and routine use of standard addition as ion selective electrodes become increasingly popular for demanding, real world applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.