Abstract

An automated test system and procedure is proposed, designed to enable systematic testing of electrical impedance tomography (EIT) devices. The system is designed to calculate reliable, repeatable and accurate performance figures of merit of an EIT system using a saline phantom and an industrial robot arm. Applications of the test system are to compare EIT devices against requirements, or to help optimize a device for its operating parameters. A test methodology and sample test results are presented to illustrate its use. The system is used to compare image quality and contrast detection for a range of stimulation and measurement patterns, and results show the best images when the pair of current injection electrodes is spaced between 45 and 170 degrees on a tank. Finally, we propose a classification of the object detection errors, which can facilitate comparison of EIT instrument specifications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.