Abstract

A procedure was developed for the extraction-photometric determination of low concentrations of mercury(II) in aqueous solutions. It is based on the reaction of the Hg Br 3 − acidocomplex with the Astraphloxin cationic dye to give an ion pair extracted with toluene. The detection limit for mercury is 0.015 mg/L. Effects of a number of heavy metal cations on determining mercury(II) were studied. It was shown that these effects appear in the presence of 500-fold molar amounts of interferents. The procedure was used for determining mercury(II) in the presence of 200-fold molar amounts of some heavy metal cations. The relative standard deviation was no worse than 5%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.