Abstract

We have used an anhepatic dog model to demonstrate the efficacy of a bioartificial liver assist device. Six dogs underwent total hepatectomy. Three received only medical care (controls) while the remainder were connected to an extracorporeal liver assist device (ELAD). The control dogs failed to regain consciousness after anesthesia although all lived 4-5 h postoperatively. Plasma ammonia concentration increased by an average of 250 mumol/L between the end of surgery and the demise of the animals. The treated dogs lived 3-12.5 h, and 2 of them required repeated doses of thiamylal sodium to maintain sedation. Plasma ammonia concentration was unchanged after connection to the ELAD except in the longest survivor, whose ammonia began to rise after 8 h on the ELAD. The short survival in the other 2 treated dogs was the result of uncontrolled intraabdominal bleeding. This device is capable of replacing the metabolic function of the liver, and might provide hepatic support in patients awaiting transplantation or in fulminant hepatic failure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.