Abstract
To achieve optimal PET image reconstruction through better system modeling, we developed a system matrix that is based on the probability density function for each line of response (LOR-PDF). The LOR-PDFs are grouped by LOR-to-detector incident angles to form a highly compact system matrix. The system matrix was implemented in the MOLAR list mode reconstruction algorithm for a small animal PET scanner. The impact of LOR-PDF on reconstructed image quality was assessed qualitatively as well as quantitatively in terms of contrast recovery coefficient (CRC) and coefficient of variance (COV), and its performance was compared with a fixed Gaussian (iso-Gaussian) line spread function. The LOR-PDFs of three coincidence signal emitting sources, (1) ideal positron emitter that emits perfect back-to-back γ rays (γγ) in air; (2) fluorine-18 (18F) nuclide in water; and (3) oxygen-15 (15O) nuclide in water, were derived, and assessed with simulated and experimental phantom data. The derived LOR-PDFs showed anisotropic and asymmetric characteristics dependent on LOR-detector angle, coincidence emitting source, and the medium, consistent with common PET physical principles. The comparison of the iso-Gaussian function and LOR-PDF showed that: (1) without positron range and acollinearity effects, the LOR-PDF achieved better or similar trade-offs of contrast recovery and noise for objects of 4 mm radius or larger, and this advantage extended to smaller objects (e.g. 2 mm radius sphere, 0.6 mm radius hot-rods) at higher iteration numbers; and (2) with positron range and acollinearity effects, the iso-Gaussian achieved similar or better resolution recovery depending on the significance of positron range effect. We conclude that the 3D LOR-PDF approach is an effective method to generate an accurate and compact system matrix. However, when used directly in expectation–maximization based list-mode iterative reconstruction algorithms such as MOLAR, its superiority is not clear. For this application, using an iso-Gaussian function in MOLAR is a simple but effective technique for PET reconstruction.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.