Abstract

Increasing urethral occlusion pressures were created with an artificial sphincter in fresh bladder-urethra specimens of adult pigs. Detrusor leakage pressures remained about, 25 per cent below the associated sphincter cuff pressures.The same relation was found when, during abrupt variations of cuff pressure, the detrusor leakage pressures were measured with antegrade urethral perfusion pressure measurement through a new sleeve catheter. Side-hole perfusion pressure measurements showed 100 per cent transmission of abrupt variations of cuff pressures to the inner urethral wall. No rotational differences could be detected.From the results it has been concluded that a probable pressure loss of about 25 per cent must be taken into account when converting artificial sphincter pressures into bladder pressures at which leakage will occur.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.