Abstract

The potential risk of inducing ventricular arrhythmias was explored by employing pulsed field ablation (PFA) through the administration of both monophasic and biphasic waveform deliveries. PFA was applied to specific locations in the right ventricle (RV, n = 5 sites) in swine (n = 2), utilizing identical settings with consistent amplitude, pulse width, and a number of pulses for both monophasic and biphasic waveforms. PFA deliveries were precisely timed in 10-ms intervals across the entire T wave. Using monophasic waveforms, ventricular fibrillation (VF) was induced 7/7 times when PFA pulses were delivered during the T wave. Biphasic waveforms resulted in no ventricular arrhythmias when PFA was delivered across the entire duration of T wave. Our findings indicate that VF can be consistently induced with monophasic pulse waveforms, but not with biphasic pulse waveforms in a porcine model.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.