Abstract

Argon cluster ion sources for sputtering and secondary ion mass spectrometry use projectiles consisting of several hundreds of atoms, accelerated to 10-20 keV, and deposit their kinetic energy within the top few nanometers of the surface. For organic materials, the sputtering yield is high removing material to similar depth. Consequently, the exposed new surface is relatively damage free. It has thus been demonstrated on model samples that it is now really possible to perform dual beam depth profiling experiments in organic materials with this new kind of ion source. Here, this possibility has been tested directly on tissue samples, 14 μm thick rat brain sections, allowing primary ion doses much larger than the so-called static secondary ion mass spectrometry (SIMS) limit and demonstrating the possibility to enhance the sensitivity of time-of-flight (TOF)-SIMS biological imaging. However, the depth analyses have also shown some variations of the chemical composition as a function of depth, particularly for cholesterol, as well as some possible matrix effects due to the presence or absence of this compound.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.