Abstract

The first undulator radiation has been extracted from the Advanced Photon Source (APS). The results from the characterization of this radiation are very satisfactory. With the undulator set at a gap of 15.8 mm (K=1.61), harmonics as high as the 17th were observed using a crystal spectrometer. The angular distribution of the third-harmonic radiation was measured, and the source was imaged using a zone plate to determine the particle beam emittance. The horizontal beam emittance was found to be 6.9±1.0 nm-rad, and the vertical emittance coupling was found to be less than 3%. The absolute spectral flux was measured over a wide range of photon energies, and it agrees remarkably well with the theoretical calculations based on the measured undulator magnetic field profile and the measured beam emittance. These results indicate that both the emittance of the electron beam and the undulator magnetic field quality exceed the original specifications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.