Abstract

We determined laser parameters and delivery system requirements for successful use of the potassium-titanyl-phosphate (KTP) laser for strabismus surgery and its histopathologic effect on extraocular muscle, tendon, sclera, and cornea of human and rabbit eyes. Enucleated cadaver eyes and exenterated rabbit orbital contents were used. Using a 200-microns fiberoptic tip delivery system, varying laser energies were used to perform muscle insertion site and corneal damage, tenotomies, and myotomies. Pulse durations of 0.5 second and energies above 0.8 watt were needed for tenotomy or myotomy. No damage to the surrounding tissues or sclera was observed with laser tenotomy or myotomy performed 0.5 mm from the insertion site. Energy above 0.8 watt applied directly caused full thickness disruption. In comparison to scissor myotomy, the laser provided charring of the superficial tissues. This study shows that the KTP laser can be safely and accurately used for extraocular muscle tenotomy or myotomy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.