Abstract

A design study of a small-size neutral particle analyzer using a silicon surface barrier detector (SSD) is performed. The SSD is very sensitive to x rays or photons, so that a pair of 45° sector magnets to separate a reionized neutral from x rays and photons will be used for this analyzer. In order to examine the performance, the SSD was applied to measure the species ratio of the prototype neutral beam injector for JT-60. It was confirmed that the energy resolution was 12% at 40 keV and the linear relation between the incident energy of particles and the pulse height was held over the energy range from 16.7 to 100 keV. The species ratio measured by the SSD was in good agreement with that by the Doppler-shift spectrometer. The SSD has a sufficient capability for the energy analysis of fast neutrals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.