Abstract

Two types of ion implanted miniature p-i-n diodes were tested in a d(48.5) + Be fast-neutron beam produced in the Detroit superconducting cyclotron. The increase in forward voltage drop caused by neutron-induced damage was correlated with neutron dose measured in a water phantom. The neutron and gamma dose components were predetermined using twin detector (Tissue-equivalent ion chamber paired with miniature Geiger-Müller counter) method. The increase in the voltage drop for 1 mA injection current was monitored together with the cyclotron beam target current, thus the differential voltage drop could be defined precisely for given radiation dose. The average neutron sensitivities of tested diodes were 1.284 +/- 0.014 and 0.528 +/- 0.058 mV per cGy. The miniature detectors can be utilised in characterisation of small radiation fields and in the regions of high dose gradient as well as for in vivo dosimetry of the patients undergoing fast-neutron therapy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.