Abstract

Low-frequency Raman (LFR) spectroscopy is presented as a viable tool for studying the hydration characteristics of lyotropic liquid crystal systems herein. Monoolein was used as a model compound, and its structural changes were probed both in situ and ex situ which enabled a comparison between different hydration states. A custom-built instrumental configuration allowed the advantages of LFR spectroscopy to be utilized for dynamic hydration analysis. On the other hand, static measurements of equilibrated systems (i.e., with varied aqueous content) showcased the structural sensitivity of LFR spectroscopy. The subtle differences not intuitively observed between similar self-assembled architectures were distinguished by chemometric analysis that directly correlated with the results from small-angle X-ray scattering (SAXS), which is the current "gold standard" method for determining the structure of such materials.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.