Abstract

This paper introduces the KRISS-Rn4, a high-sensitivity radon monitor with four detection cells, installed within a walk-in type radon calibration chamber at KRISS. The KRISS-Rn4 exhibits enhanced energy resolution through channel-by-channel signal processing and data acquisition. Results reveal that it achieves lower statistical fluctuations and faster response times in monitoring test atmospheres compared to commercial devices.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.