Abstract
An investigation was performed in order to establish if dried blood spots (DBS) could be applied to the quantitation of biopharmaceuticals in biological matrices and perform equivalently in terms of accuracy, precision and stability to traditional plasma methods. A method was successfully validated for the peptide Exendin-4 (39 amino acids in length) utilizing DBS technology. The validated DBS method resulted in a more sensitive and simplistic method than an existing monkey plasma method and required tenfold less sample volume. The final DBS method resulted in a 10-2000-ng/ml linear calibration range using approximately 5 µl of dried blood, compared with the plasma method in which 150 µl of plasma coupled with SPE sample preparation resulted in a 20-2000-ng/ml linear calibration range. Although not needed for DBS, SPE was required for the plasma method to reduce endogenous matrix interferences and achieve desired LLOQ. Matrix stability was also enhanced by the implementation of the DBS platform when compared with either plasma or whole blood. DBS technology can be utilized for the quantitation of biopharmaceuticals and offer advantages over traditional plasma-based methods.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.