Abstract
A novel type of adiabatic scanning calorimeter (ASC) based on Peltier elements (PEs) is used to obtain high-resolution enthalpy and heat capacity data on the melting transition of gallium. The accuracy of the specific heat capacity and specific enthalpy is about 2 %, for a sub-mK temperature resolution. The simultaneously determined equilibrium specific heat capacity and specific enthalpy are used to determine the heat of fusion and the purity. In addition, the use of the PE-based ASC as a classical heat step calorimeter and as a constant rate (DSC-type) calorimeter is discussed. A comparison of the ASC results with literature data and DSC data shows the advantages of ASC for the study of phase transitions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.