Abstract

Microwave ablation is a technique used in treating hepatocellular carcinoma, especially in cases where surgical removal is impossible. In the present study we are investigating the effects of design characteristics of a coaxial slot antenna (single- versus double-slot, slot-to-tip distance and slot size) on the ablation zone characteristics (dimensions and shape). The specific absorption rate field and the temperature rises are calculated for a variety of application times and powers. A plateau in the ablation zone dimensions in healthy and cirrhotic liver models is predicted, but not in malignant ones. It is found that the value of the perfusion rate (which is different for each clinical case) is of crucial importance in order to correctly estimate the ablation zone. An underestimation of dimensions is expected, if higher perfusion rates are used (i.e., values for healthy tissue rather than malignant). In contrast, an exact determination of the values of relative permittivity and conductivity is less significant for predicting the ablation zone.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.