Abstract

Detailed surgical management, magnetic resonance imaging (MRI), and computer tomography (CT) images of a broken annular closure device (ACD) have not been reported yet. In this case, a 28-year-old male presented with a new onset of radiculopathy three years after lumbar discectomy and placement of an ACD. The CT-myelography and MRI revealed a recurrent disc herniation (RDH) and dislocation of a broken ACD. ACD removal was performed and confirmed breakage due to RDH with scarring around the RDH and displaced ACD. Implant-associated complications and management should be reported in detail in order to enhance knowledge on device-related complications.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.