Abstract

The goals of this investigation were to measure the anisotropy of backscattered ultrasound observed in the short-axis view of mouse hearts in systole and diastole and to compare these measurements with predictions from a computer simulation. Measurements of midmyocardial apparent backscatter were obtained from analyses of the hearts of seven wild-type mice using a clinical imaging system utilizing a linear array with a nominal center frequency of 13 MHz. A computer model simulating the short-axis view was implemented based on previous measurements of the angle of insonification dependence of myocardial backscatter and attenuation. Results demonstrate that the measured backscatter was largest for those myocardial regions corresponding to approximately perpendicular insonification relative to the myofibers and the smallest for regions of approximately parallel insonification, with the minimum to maximum values of apparent backscatter differing by approximately 10 dB. The measured anisotropy of backscatter was similar for end-systole and end-diastole and was in good agreement with the predicted anisotropy obtained from the computer simulations. (E-mail: mrh@wuphys.wustl.edu)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.