Abstract

A prototype directional atherectomy catheter (Omnicath) was evaluated in four Yucatan microswines. Atherectomy was performed on iliac or aortic target lesions. After control angiography, the animals were sacrificed and the target arteries were examined histologically. Atherectomy resulted in arterial ruptures in three cases, and the track of the blade was measured to be of an average depth of 0.38 mm. Maneuverability was satisfactory but aspiration was not efficient. Precise localization of the atherectomy window was difficult. We conclude that modification of the catheter seems mandatory before use in humans.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.