Abstract

Objective: This study developed a sensitive, selective, and valid method for analyzing clopidogrel in human plasma using liquid chromatographytandemmass spectrometry (LC-MS/MS).Methods: The chromatography separation was performed on the waters ACQUITY UPLC Class BEH C18 1.7 μm (2.1 × 100 mm) column, consisting of0.1% formic acid in water and 0.1% formic acid in acetonitrile (30-70) as the mobile phase, and an isocratic elution with a flow rate of 0.2 mL/minutes.The mass detection was performed using a Waters Xevo TQD equipped with positive electrospray ionization in the multiple reaction monitoringmodes. The clopidogrel was detected at m/z 322.086>212.097 and irbesartan as an internal standard at m/z 429.233>207.131. The sample wasprepared with protein precipitation method using acetonitrile, vortex mixed for 10 minutes, and centrifuged at 13,000 rpm for 20 minutes.Results: This method showed a linear result at the concentration range of 0.2-10 ng/ml with r≥0.9997.Conclusions: The developed method provides sensitivity, linearity, precision, and accuracy and is suitable for analysis of clopidogrel in human plasmausing LC-MS/MS.

Highlights

  • Clopidogrel is used for the secondary prevention of cerebral or myocardial infarction in patients who cannot tolerate acetylsalicylic acid or those who undergo a cerebral or myocardial infarction while taking acetylsalicylic acid [1]

  • Clopidogrel is a prodrug, with a slow onset, that has been found to be as effective as acetylsalicylic acid for secondary prevention in stroke

  • In 2012, another study conducted a validation of the clopidogrel analytical method through protein precipitation, which was simpler and did not use methanol in the analysis [8]

Read more

Summary

Methods

The chromatography separation was performed on the waters ACQUITY UPLC Class BEH C18 1.7 μm (2.1 × 100 mm) column, consisting of 0.1% formic acid in water and 0.1% formic acid in acetonitrile (30-70) as the mobile phase, and an isocratic elution with a flow rate of 0.2 mL/minutes. The mass detection was performed using a Waters Xevo TQD equipped with positive electrospray ionization in the multiple reaction monitoring modes. The clopidogrel was detected at m/z 322.086>212.097 and irbesartan as an internal standard at m/z 429.233>207.131. The sample was prepared with protein precipitation method using acetonitrile, vortex mixed for 10 minutes, and centrifuged at 13,000 rpm for 20 minutes

Conclusions
INTRODUCTION
METHODS
RESULTS AND DISCUSSION
CONCLUSION

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.