Abstract
A method is developed for calculating the time dispersion of amplitude and risetime compensated (ARC) timing with germanium planar detectors. Universal curves for pulse shape dependent timing error and for time jitter indicate that there is an optimum ratio between the inverted and delayed pulses which leads to the best realization of a given experimental objective. Curves for the corresponding system parameters are given. It is shown that ARC timing with Ge planar detectors is always capable of giving superior performance than constant fraction timing. The work includes quantitative comparison with experimental results published elsewhere.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.