Abstract

The effects of time-varying readout gradients and data sampling with variable dwell times in magnetic resonance imaging are examined. General reconstruction formulas are given for linear reconstruction with even k-space weighting. Closed analytic expressions for estimator variance are given for data sampling during arbitrary gradient waveforms with both uniform kx step size and nonuniform kx step size. It is shown that estimator variance increases (the signal-to-noise ratio decreases) for nonconstant gradient waveforms. It is also shown that estimator variance is greater for constant k-space sampling strategies than for constant time sampling at the Nyquist rate. Data collected during a triangular readout gradient waveform, with either constant time or constant k-space sampling, versus conventional (constant gradient) collection confirms theoretical predictions for estimator variance. The benefits of collecting data while the readout gradient is ramping up from and down to zero are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.