Abstract

The present study utilized Analytical Quality by Design (AQbD) approach to develop a stability-indicating high-performance liquid chromatography (HPLC) method for estimating evogliptin tartrate using design expert software. The key parameters were methodically optimized, contours were plotted, and stability was evaluated using various forced degradation conditions. Using an Agilent HPLC system with a photo diode array (PDA) detector along with Fortis C18 column (250 × 4.6mm, 5μm) effectively separated the drug from its degradants. The mobile phase used was methanol: water (pH adjusted to 3.0, 76:24; v/v) at 0.8mL/min flow rate. Evogliptin was eluted at 2.98 min, at a detection wavelength of 267 nm. The proposed method was found to be specific, precise, linear and robust. The drug was sensitive to acidic, basic, oxidative, thermal, and photodegradation resolving six degradation products. Thus, the developed AQbD-based stability-indicating HPLC method is applicable in analyzing evogliptin in bulk, tablet dosage form and stability samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.