Abstract

Automated instruments that can examine urine for cells and particles have reduced the need for labor-intensive manual microscopy. We evaluated the performance of the Iris iQ200 Automated Urine Microscopy Analyzer (Iris Diagnostics, Chatsworth, CA) and compared results with manual cell and particle counts using Fuchs-Rosenthal counting chambers. Within-run imprecision (coefficient of variation) of the iQ200 for urine samples ranged from 3.0% to 45% for RBC counts between 1,029 and 3 x 10(6) cells/L, 3.4% to 40% for WBC counts between 1,006 and 4 x 10(6) cells/L, and 8.9% to 35% for epithelial cell counts between 93 and 4 x 10(6) cells/L. Between-run imprecision was 3.3% at 1,017 x 10(6) cells/L and 19.2% at 28 x 10(6) cells/L. There was good agreement between the iQ200 and manual cell counts (r > or = 0.94); however; the iQ200 produced lower results based on slopes of 0.92 (RBC count), 0.81 (WBC count), and 0.94 (epithelial cell counts). The iQ200 has satisfactory performance and correlates well with manual cell counts. Most urine samples containing RBCs, WBCs, and epithelial cells can be reported without review of captured images.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.