Abstract

A method using purge-and-membrane mass spectrometry (PAM-MS) was developed for the analysis of residual solvents in pharmaceutical products. The method combines dynamic headspace and membrane inlet mass spectrometry. The limits of detection for the compounds studied, benzene, toluene, chloroform, 2-pentene and 2-methyl- and 3-methylpentane, were 0.05-0.1 mg/kg. In quantitative analysis the method showed good linearity (r(2) > 0.998) and acceptable within-day (RSD = 7.9-18%) and between-day (RSD = 6.8-10%) repeatability. The PAM-MS method combined with the custom-made Solver program was compared with a method using purge-and-trap gas chromatography/mass spectrometry (P&T-GC/MS) for identification of residual solvents from authentic samples. The results showed that PAM-MS/Solver provides reliable identification of the main volatile organic compounds (VOCs) in the pharmaceuticals, but VOCs with low concentrations (below 0.5 mg/kg) were better identified by P&T-GC/MS. Other advantages of the PAM-MS method were short analysis times and non-requirement for pre-treatment of samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.