Abstract

To report how peak inflating pressure (PIP), tidal volume (VT) and low-tidal volume alarms are affected by maximum allowed inflating pressure (Pmax) during volume guarantee (VG) ventilation. Ventilation data were analysed with 1 Hz sampling rate from 25 neonates receiving synchronised intermittent positive pressure ventilation with VG for >12 h. The difference between Pmax and PIP (Pdiff) ranged between 5 and 20 mbar (median = 11 mbar) despite a protocol to keep Pmax 5 mbar above the "working PIP". Pmax was reached in 5.2% of inflations. Computational modelling demonstrated that had Pdiff been kept at 5 mbar more consistently, >10% of inflations would have reached Pmax. The frequency of low-tidal volume alarms showed inverse correlation with Pdiff. It is difficult to implement a simple Pmax strategy due to variability of PIP. Setting Pmax close to the "working PIP" limits VT delivery and triggers frequent alarms.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.