Abstract

We analyze the feasibility of using refractive free-form phase plates at the aperture stop of microscope objectives as an alternative to active alignment to compensate for assembly tolerances. The method involves the determination of misalignment-induced aberrations at the exit pupil, and transferring them to the aperture stop while taking pupil aberrations into consideration. We demonstrate that despite being able to correct only for field-independent aberrations, this method can restore near-diffraction-limited imaging performance of passively aligned systems with practical tolerances, given that the as-designed system is highly corrected. We confirm the results via numerical simulations for two different commercial objective designs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.