Abstract
Identifying the elbow flexion-extension (F-E) movement axis is important for placing a hinged elbow external fixator. An X-ray fluoroscopy-based method is widely used in clinical practice, exposing the patient and surgeons to high doses of radiation. Additionally, the accuracy and repeatability of the fluoroscopy-based method are very low and affected by subjective factors. To solve this problem, an X-ray-free method based on kinematics analysis was proposed to identify the elbow F-E movement axis, and a navigation system was built to guide the placement of the elbow external fixator. Our X-ray-free navigation method is more repeatable than the current X-ray fluoroscopy method used clinically. Both our algorithm and the NIST (National Institute of Standards and Technology) algorithm showed high accuracy and repeatability to identify the axis. The method proposed in this study is very promising to avoid a large dose of X-ray radiation and increases the repeatability and performance of identifying the elbow F-E movement axis for the placement of the hinged elbow external fixator.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International journal of computer assisted radiology and surgery
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.