Abstract In Kjeldahl methods for the determination of nitrogen in fertilizers, the digestion temperature must be above 370°C to mineralize the nitrogen in refractory organic compounds such as nicotinic acid. In total nitrogen determinations by the Raney catalyst powder method, temperatures during the digestion step were investigated as a possible cause of occasional low results. Decomposition of the polyethylene envelopes of Kel-Paks in the presence of salts of metallic reductants caused rapid fluctuations and excessive rise in temperature with consequent loss of ammoniacal nitrogen. Complete Kel-Paks can be used without difficulty in methods such as 2.044 (for nitrate-free samples) or 2.045 (salicylic acid) that do not use metallic reductants, but only the contents of the Kel-Paks without the envelopes should be used when Raney catalyst or chromium powders have been used to reduce nitrates.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.