Abstract

In this study we evaluated the minimum stimulating current associated with intraneural needle placement and sonographic appearance of intraneural injection. We inserted a needle 2 cm inside 28 pig nerves (brachial plexus in vivo), recorded the minimum current to elicit a motor response, and injected dye (5 mL) under ultrasound (US) imaging. The minimum current to elicit a motor response was 0.43 mA (range: 0.12-1.8 mA). Nerve expansion was visualized by US in 24 of 28 nerves. Histology revealed penetration of the epineurium in these same 24 nerves. There was no evidence of dysplasia within the fascicle of any nerve. US may prove useful to detect intraneural injection, whereas a motor response above 0.5 mA may not exclude intraneural needle placement. The correlation between intraneural injection and neurological dysfunction remains unclear.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.