Abstract

The paper describes the theory and application of opto-acoustics to determine thermal diffusivities of gases. An experimental device, already described in previous papers of the authors [Internat. J. Thermophys. 19 (1998) 1099; Proc. 2nd European Thermal Science and 14th UIT National Heat Transfer Conf., 1996, pp. 1071–1078] permitted the detection of thermal diffusivities of gases at moderate pressures with an experimental uncertainty of about ±1.25%.Based on the experience gained with this device, a comprehensive error analysis is presented in this paper. It shows how the experimental uncertainties can be considerably reduced to about −0.45 to +0.35%. The parameters for optical cell design are dealt with, as well as the appropriate characteristics, such as frequencies of the modulated laser beam, and the microphone used in the experiment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.