Abstract

We have investigated the extent to which antireflux (non-return) valves impede the maximum flow rate of crystalloid in intravenous infusion systems when administered either under gravity or with 300 mmHg added pressure. When used with a 16G cannula, a reduction in flow rate of 19-38% was consistently seen. With a 20G cannula, no significant reduction in flow was found. The valve tested that produced the least reduction in flow was the Westcott Sae-flo valve.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.