Abstract

Measurements of central ray tissue-air ratio (TAR) for 60Co for a large number of irregularly shaped fields typically encountered in the cancer clinic were compared with TAR values either calculated by the Clarkson method using scatter air ratios or obtained from tables of square field data using the area/perimeter approach. Irregular field shapes included the "L", pentagonal, rectangular, upper and lower mantle and split fields. Agreement between both calculational methods and direct measurement is within +/- 2% if careful attention is given to central ray position as affected by the proper use of off-centre ratios in air. In the case of 10 MV X-rays, eight randomly chosen irregular fields again yielded an average agreement between both methods of calculation and experimental data of less than 1% and an agreement of less than 0.5% between the A/P and Clarkson calculations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.