Abstract

An analytical dose calculation algorithm was developed and commissioned to calculate dose delivered with both static and dynamic multileaf collimator (MLC) in a homogenous phantom. The algorithm is general; however, it was designed specifically to accurately model dose for large and complex IMRT fields. For such fields the delivered dose may have a considerable contribution from MLC transmission, which is dependent upon spatial considerations. Specifically, the algorithm models different MLC effects, such as interleaf transmission, the tongue-and-groove effect, rounded leaf ends, MLC scatter, beam hardening and divergence of the beam, which results in a gradual MLC transmission fall-off with increasing off-axis distance. The calculated dose distributions were compared to measured dose using different methods (film, ionization chamber array, single ionization chamber), and the differences among the treatment planning system, the measurements and the developed algorithm were analysed for static MLC and dynamic IMRT fields. It was found that the calculated dose from the developed algorithm agrees very well with the measurements (mostly within 1.5%) and that a constant value for MLC transmission is insufficient to accurately predict dose for large targets and complex IMRT plans with many monitor units.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.