Abstract

Fluid pressure may stimulate osteolysis near screw holes in joint arthroplasty components. We developed a generalized in vitro model of a polyethylene liner and metal backing with a screw hole to investigate whether implant design factors influence local fluid pressure. We observed an order of magnitude of variation in the peak screw hole pressure (from 16.0 and 163 kPa) under clinically relevant loading conditions. Of the implant factors investigated, the surface finish of the metallic base plate had the greatest effect on peak screw hole fluid pressures; the thickness of the polyethylene liner, as well as the gap between the liner and the base plate, were also significant design variables. Our data suggest that unpolished metal base plates, thick polyethylene liners, and tight conformity between the liner and the metal base plate will all contribute to significantly reduced peak screw hole fluid pressures in joint arthroplasty.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.