Abstract

An HPLC method for digoxin quantification in dissolution samples obtained as per the official British Pharmacopeia (BP) method is presented in this paper. The chromatography was performed at 20 ?C on a Symmetry C18; 3.5 ?m, 75 x 4.6 mm column with water - acetonitrile (72 : 28, v/v), as the mobile phase and UV detection at 220 nm. The method was found to be selective, linear, accurate and precise in the specified ranges. The LOD and LOQ were 0.015 ?g mL-1 and 0.050 ?g mL-1, respectively. Robustness testing was conducted to evaluate the impact of minor changes in the chromatographic parameters (i.e., acetonitrile fraction, flow rate of the mobile phase, column temperature and column length) on the characteristics of the digoxin peak. A. full factorial design (24) was used to investigate the influence of the four variables The presented HPLC method was applied in quality and stability testing of Digoxin tablets 0.25 mg.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.