Abstract

To evaluate the effects of eyeball structure on ocular pulse amplitude (OPA) measured using dynamic contour tonometer (DCT). In 86 eyes of 43 healthy subjects, we measured OPA and intraocular pressure (IOP) with DCT (DCT-IOP), IOP with Goldmann applanation tonometry (GAT-IOP), central corneal thickness (CCT), corneal thickness 2 mm (2 mmCT) and 4 mm (4 mmCT) apart from the center, corneal volume within a 3.5-mm radius from the corneal center, corneal curvature, anterior chamber depth, anterior chamber volume, and axial length (AL). OPA had a significant positive correlation with GAT-IOP (Pearson's r = 0.412, p < 0.001), DCT-IOP (r = 0.350, p < 0.001), and 4 mmCT (r = 0.244, p = 0.0231), and had a significant negative correlation with AL (r = -0.268, p = 0.0122). In a multiple linear regression analysis, AL and GAT-IOP were significantly associated with OPA. OPA measured with DCT is significantly influenced by several factors, such as IOP, peripheral corneal thickness (4 mmCT), and AL.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.