Abstract

External defibrillation is a common treatment for the cardiac arrhythmia atrial fibrillation. Electrode placement has been shown to affect defibrillation efficacy and required energy levels. We suggest investigating the relationship between esophageal electric fields (EEFs) and atrial defibrillation thresholds to determine the feasibility of creating patient-specific electrode placements using EEFs. This study presents the design and implementation of an esophageal probe (EP) that accurately measures three-dimensional electric fields. The root-mean-square error of the EP was 1.69% as determined by measurements performed in an electrolytic tank. The EP also performed well during in vivo testing in a pig. There was a strong positive relationship between EEF2s and applied energy during defibrillation strength shocks. The EEF measurements were also repeatable, with less than 4.24% difference between repeated shocks. This is the first description of a probe designed specifically for measuring electric fields in the esophagus.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.