Abstract

An empirical analytical function has been obtained for the full energy peak efficiency of a co-axial P-type High Purity Germanium Detector for gamma-ray energies in the range of 80 - 1333 keV and source-to-detector distance in the range of 1.0 - 4.0 cm. Comparison of the calculated efficiencies with the corresponding experimental value agrees very well within standard deviations of 1.8 - 3.7 %. JOURNAL OF THE GHANA SCIENCE ASSOCIATION Volume 2 No. 1 (2000) pp. 64-70

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.