Abstract

An automated flow calorimeter has been developed for the measurement of heat capacity and latent enthalpies of fluids at elevated temperatures (300–700 K) and pressure (<30M Pa) with a design accuracy of 0.1%. The method of measurement is the traditional electrical power input flow calorimeter, utilizing a precision metering pump, which eliminates the need for flow-rate monitoring. The calorimeter cell uses a unique concentric coil design with passive metal radiation shields and active guard heaters to minimize heat leakage, eliminate the traditional constant-temperature bath, and facilitate easy component replacement. An additional feature of the instrument is a complete automation system, greatly simplifying operation of the apparatus. A novel multitasking software scheme allows a single microcomputer simultaneously to control all system temperatures, provide continuous monitoring and updates on system status, and log data. Preliminary results for liquid water mean heat capacities show the equipment to be performing satisfactorily, with data accuracies of better than ±0.3%. Minor equipment modifications and better thermometry are required to reduce systemic errors and to achieve the designed operational range.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.