Abstract
An open tubular (OT) sample preparation/separation platform was developed. A multi-channel polymer layer open tubular (mPLOT) solid phase extraction (SPE) column was prepared by wall-coating the 126 channels (8μm inner diameter (ID) each) of a crystal fiber capillary with an organic polymer, namely poly(styrene-co-octadecene-co-divinylbenzene) (PS-OD-DVB). The mPLOT SPE was coupled on-line with a 10μm×2m poly(styrene-co-divinylbenzene) (PS-DVB) OT liquid chromatography column with nanospray mass spectrometry (OTLC–MS). Compared to using monolithic/particle-packed SPEs, mPLOT-SPE-OTLC allowed both fast loading and sufficient refocusing on the OT analytical column of small model compounds (sulfonamides≈300Da). Using automatic filtration/filter back-flushing (AFFL) plumbing, the mPLOT SPE column gave a constant and low back-pressure ≈35bar at 0.5μL/min. Surprisingly large sample volumes (10μL) were possible to be injected using a 12cm mPLOT.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.